Polarization-sensitive optical coherence tomography measurements with different phase modulation amplitude when using continuous polarization modulation

Lu, Zenghai, Kasaragod, Deepa K. and Matcher, Stephen J. (2012) Polarization-sensitive optical coherence tomography measurements with different phase modulation amplitude when using continuous polarization modulation.

Abstract

We demonstrate theoretically and experimentally that the phase retardance and relative optic-axis orientation of a sample can be calculated without prior knowledge of the actual value of the phase modulation amplitude when using a polarization-sensitive optical coherence tomography system based on continuous polarization modulation (CPM-PS-OCT). We also demonstrate that the sample Jones matrix can be calculated at any values of the phase modulation amplitude in a reasonable range depending on the system effective signal-to-noise ratio. This has fundamental importance for the development of clinical systems by simplifying the polarization modulator drive instrumentation and eliminating its calibration procedure. This was validated on measurements of a three-quarter waveplate and an equine tendon sample by a fiber-based swept-source CPM-PS-OCT system.

Information
Library
View Item